HZ Series
- Designed to provide good attenuation to RFI noise in the frequency range from 10kHz to 30MHz
- Size and cost-effective
- Low leakage current
- New versions up to 30A
Ordering Information

Available Part Numbers
Click on any part number for 3D CAD models, customer drawings, distributor stock status and eCommerce Links.
|
|
Specifications
| Maximum leakage current, each Line to Ground: |
| |
@ 120 VAC 60 Hz |
2 µA |
| |
@ 250 VAC 50 Hz |
5 µA |
| |
|
|
| Hipot rating (one minute): |
| |
Line to Ground |
2250 VDC |
| |
Line to Line |
1450 VDC |
| |
|
|
| Operating Frequency: |
50 / 60 Hz |
| |
|
|
| Rated voltage (max.): |
250 VAC |
| |
|
|
| Rated Current: |
3A |
| |
|
| Operating Ambient Temperature Range (at rated current Ir): |
| |
|
-10°C to +40°C |
In an ambient temperature (Ta) higher than +40°C
the maximum operating current (Io) is calculated as
follows: Io = Ir √(85-Ta)/45
|
Electrical Schematic

|
| Case Styles |
3EHZ1

| Typical Dimensions: |
| |
Line/Load Terminals(4): |
.250 [6.3] with .07 [1.8] Dia. hole |
| |
Ground Terminal(1): |
.250 [6.3] with .07 x .16 [1.8 x 3.8] slot |
| |
Mounting Holes (2): |
.188 [4.75] Dia. |
4EHZ1

| Typical Dimensions: |
| |
Line/Load Terminals(4): |
.250 [6.3] with .07 [1.8] Dia. hole |
| |
Ground Terminal(1): |
.250 [6.3] with .07 x .16 [1.8 x 3.8] slot |
| |
Mounting Holes (2): |
.188 [4.75] Dia. |
6HZ1

| Typical Dimensions: |
| |
Line/Load Terminals(4): |
.250 [6.3] with .07 [1.8] Dia. hole |
| |
Ground Terminal(1): |
.250 [6.3] with .07 x .16 [1.8 x 3.8] slot |
| |
Mounting Holes (2): |
.188 [4.75] Dia. |
10, 15 & 20EHZ1

| Typical Dimensions: |
| |
Line/Load Terminals(4): |
.250 [6.3] with .07 [1.8] Dia. hole |
| |
Ground Terminal(1): |
.250 [6.3] with .07 x .16 [1.8 x 3.8] slot |
| |
Mounting Holes (2): |
.188 [4.75] Dia. |
|
|
30EHZ6

| Typical Dimensions: |
| |
Terminals(5): |
8-32, Torque 18 lbf-in. [2.03 N-m] max ±2 [.22] |
| |
Mounting Holes (2): |
.188 [4.75] Dia. |
Case Dimensions
| Part No. |
A |
B |
C |
D |
E |
| (max.) |
(max.) |
(max.) |
±.015 |
(max.) |
| ± .38 |
| 3EHZ1 |
3.54 |
2.08 |
1.31 |
2.938 |
3.35 |
| 89.9 |
52.8 |
33.3 |
74.63 |
85.1 |
|
| 4EHZ1 |
3.07 |
1.82 |
1.16 |
2.375 |
2.78 |
| 77.98 |
46.23 |
29.46 |
60.33 |
70.61 |
|
| 6EHZ1 |
3.07 |
1.82 |
1.28 |
2.375 |
2.78 |
| 77.98 |
46.23 |
32.51 |
60.33 |
70.61 |
|
10EHZ1
15EHZ1
20EHZ1 |
3.54 |
2.047 |
1.805 |
2.938 |
3.54 |
| 89.92 |
51.99 |
45.85 |
74.63 |
89.92 |
|
| 30EHZ6 |
4.92 |
2.07 |
1.53 |
3.947 |
4.33 |
| 124.97 |
52.58 |
38.86 |
100.25 |
109.98 |
|
|
Performance Data
| Typical Insertion Loss |
| Measured in closed 50 ohm system |
| |
3EHZ1

|
4EHZ1

|
6EHZ1

|
10EHZ1

|
15EHZ1

|
20EHZ1

|
30EHZ6

|
|
 |
| Minimum Insertion Loss in dB: |
| Measured in closed 50 ohm system |
| |
|
|
| Common Mode / Asymmetrical (Line to Ground) |
|
Differential Mode / Symmetrical (Line to Line) |
|
|
|
| Model |
Frequency - MHz |
| .01 |
.05 |
.1 |
.15 |
.5 |
1 |
5 |
10 |
30 |
|
| 3EHZ1 |
10 |
24 |
30 |
34 |
34 |
35 |
15 |
10 |
5 |
| 4EHZ1 |
12 |
24 |
31 |
35 |
47 |
47 |
30 |
25 |
18 |
| 6EHZ1 |
9 |
21 |
27 |
30 |
36 |
34 |
27 |
22 |
16 |
| 10EHZ1 |
7 |
21 |
25 |
31 |
43 |
40 |
26 |
21 |
14 |
| 15EHZ1 |
7 |
27 |
27 |
30 |
43 |
37 |
24 |
17 |
12 |
| 20EHZ1 |
5 |
19 |
24 |
28 |
31 |
29 |
14 |
9 |
4 |
| 30EHZ6 |
- |
5 |
11 |
14 |
27 |
30 |
20 |
17 |
14 |
|
|
|
| Model |
Frequency - MHz |
| .01 |
.05 |
.1 |
.15 |
.5 |
1 |
5 |
10 |
30 |
|
| 3EHZ1 |
10 |
25 |
30 |
54 |
70 |
70 |
65 |
55 |
55 |
| 4EHZ1 |
- |
- |
14 |
32 |
72 |
83 |
68 |
63 |
30 |
| 6EHZ1 |
- |
- |
7 |
17 |
59 |
80 |
67 |
60 |
52 |
| 10EHZ1 |
- |
- |
4 |
21 |
63 |
80 |
80 |
74 |
36 |
| 15EHZ1 |
- |
- |
7 |
15 |
51 |
77 |
80 |
74 |
48 |
| 20EHZ1 |
- |
- |
11 |
9 |
54 |
77 |
74 |
69 |
47 |
| 30EHZ6 |
- |
- |
13 |
14 |
47 |
67 |
76 |
70 |
58 |
|
|
|
| USA Customer Service: |
1-800-468-2023 |
| Corcom Products Direct: |
1-847-680-7400 |
| Germany: |
49-89-6089-0 |
|
|
Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change. |