Skip to content
Home > Series > Medical > MV

MV Series

Multi-purpose Medical Filter for Power Line Noise Protection


UL Recognized
CSA Certified
VDE Approved

 

Catalog Data Sheet
Printer friendly version

 

MV Series

  • Multi-purpose medical filter
  • Improved Line to Ground performance
  • A good solution to emission or immunity problems
  • Meets leakage current requirements of UL2601 for health care equipment

Ordering Information

Available Part Numbers

Click on any part number for 3D CAD models, customer drawings, distributor stock status and eCommerce Links.

3MV1 6MV1
10MV1 20MV1

 

 

Specifications

Maximum leakage current, each Line to Ground:
  @ 120 VAC 60 Hz .07 mA
  @ 250 VAC 50 Hz .13 mA
     
Hipot rating (one minute):
  Line to Ground 2250 VDC
  Line to Line 1450 VDC
     
Operating Frequency: 50 / 60 Hz
     
Rated voltage (max.): 250 VAC
     
Rated Current: 3A to 20A
   
Operating Ambient Temperature Range (at rated current Ir):
    -10°C to +40°C
In an ambient temperature (Ta) higher than +40°C
the maximum operating current (Io) is calculated as
follows: Io = Ir √(85-Ta)/45
   

Electrical Schematic

Case Styles

MV1 (3, 6, 10A)

Typical Dimensions:
  Line/Load Terminals(4): .250 [6.3] with .07 [1.8] Dia. hole
  Ground Terminal(1): .250 [6.3] with .07 x .16 [1.8 x 3.8] slot
  Mounting Holes (2): .188 [4.75] Dia.

20MV1

Typical Dimensions:
  Line/Load Terminals(4): .250 [6.3] with .07 [1.8] Dia. hole
  Ground Terminal(1): .250 [6.3] with .07 x .16 [1.8 x 3.8] slot
  Mounting Holes (2): .188 [4.75] Dia.

 

 

Case Dimensions

Part No. A B C D E
(max.) (max.) (max.) ±.015 (max.)
± .38
3MV1 3.36 1.82 1.28 2.375 2.78
85.3 46.2 32.5 60.33 70.6

6MV1, 10MV1 3.86 2.08 1.53 2.938 3.34
98.0 52.8 38.9 74.63 84.8

20MV1 5.23 3.38 1.53 3.75 4.20
132.8 85.9 38.9 95.25 106.7

 

Performance Data

Typical Insertion Loss
Measured in closed 50 ohm system
 

3MV

6MV

10MV

20MV

Minimum Insertion Loss in dB:
Measured in closed 50 ohm system
     
Common Mode / Asymmetrical (Line to Ground)   Differential Mode / Symmetrical (Line to Line)

 
Current Rating Frequency - MHz
.15 .5 1 2 5 10 20 30

3A 14 19 20 30 46 40 34 31
6A 19 27 30 38 50 40 35 35
10A 15 25 26 34 46 50 44 42
20A 18 30 34 34 46 40 36 20

 
Current Rating Frequency - MHz
.15 .5 1 2 5 10 20 30

3A 33 65 65 65 65 60 53 50
6A 40 65 65 65 65 60 57 55
10A 33 65 65 65 65 65 55 55
20A 25 65 65 65 65 60 57 45

 

USA Customer Service: 1-800-468-2023
Corcom Products Direct: 1-847-680-7400
Germany: 49-89-6089-0
  Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change.