| Multi-purpose Medical Filter for Power Line Noise Protection |
  
UL Recognized
CSA Certified
VDE Approved
Catalog Data Sheet 
Printer friendly version |
|
 |
MV Series
- Multi-purpose medical filter
- Improved Line to Ground performance
- A good solution to emission or immunity problems
- Meets leakage current requirements of UL2601 for health care equipment
Ordering Information

Available Part Numbers
Click on any part number for 3D CAD models, customer drawings, distributor stock status and eCommerce Links.
|
|
Specifications
| Maximum leakage current, each Line to Ground: |
| |
@ 120 VAC 60 Hz |
.07 mA |
| |
@ 250 VAC 50 Hz |
.13 mA |
| |
|
|
| Hipot rating (one minute): |
| |
Line to Ground |
2250 VDC |
| |
Line to Line |
1450 VDC |
| |
|
|
| Operating Frequency: |
50 / 60 Hz |
| |
|
|
| Rated voltage (max.): |
250 VAC |
| |
|
|
| Rated Current: |
3A to 20A |
| |
|
| Operating Ambient Temperature Range (at rated current Ir): |
| |
|
-10°C to +40°C |
In an ambient temperature (Ta) higher than +40°C
the maximum operating current (Io) is calculated as
follows: Io = Ir √(85-Ta)/45
|
| |
|
Electrical Schematic

|
Case Styles
MV1 (3, 6, 10A)

| Typical Dimensions: |
| |
Line/Load Terminals(4): |
.250 [6.3] with .07 [1.8] Dia. hole |
| |
Ground Terminal(1): |
.250 [6.3] with .07 x .16 [1.8 x 3.8] slot |
| |
Mounting Holes (2): |
.188 [4.75] Dia. |
20MV1

| Typical Dimensions: |
| |
Line/Load Terminals(4): |
.250 [6.3] with .07 [1.8] Dia. hole |
| |
Ground Terminal(1): |
.250 [6.3] with .07 x .16 [1.8 x 3.8] slot |
| |
Mounting Holes (2): |
.188 [4.75] Dia. |
|
|
Case Dimensions
| Part No. |
A |
B |
C |
D |
E |
| (max.) |
(max.) |
(max.) |
±.015 |
(max.) |
| ± .38 |
| 3MV1 |
3.36 |
1.82 |
1.28 |
2.375 |
2.78 |
| 85.3 |
46.2 |
32.5 |
60.33 |
70.6 |
|
| 6MV1, 10MV1 |
3.86 |
2.08 |
1.53 |
2.938 |
3.34 |
| 98.0 |
52.8 |
38.9 |
74.63 |
84.8 |
|
| 20MV1 |
5.23 |
3.38 |
1.53 |
3.75 |
4.20 |
| 132.8 |
85.9 |
38.9 |
95.25 |
106.7 |
|
|
Performance Data
| Typical Insertion Loss |
| Measured in closed 50 ohm system |
| |
3MV

|
6MV

|
10MV

|
20MV

|
 |
| Minimum Insertion Loss in dB: |
| Measured in closed 50 ohm system |
| |
|
|
| Common Mode / Asymmetrical (Line to Ground) |
|
Differential Mode / Symmetrical (Line to Line) |
|
|
|
| Current Rating |
Frequency - MHz |
| .15 |
.5 |
1 |
2 |
5 |
10 |
20 |
30 |
|
| 3A |
14 |
19 |
20 |
30 |
46 |
40 |
34 |
31 |
| 6A |
19 |
27 |
30 |
38 |
50 |
40 |
35 |
35 |
| 10A |
15 |
25 |
26 |
34 |
46 |
50 |
44 |
42 |
| 20A |
18 |
30 |
34 |
34 |
46 |
40 |
36 |
20 |
|
|
|
| Current Rating |
Frequency - MHz |
| .15 |
.5 |
1 |
2 |
5 |
10 |
20 |
30 |
|
| 3A |
33 |
65 |
65 |
65 |
65 |
60 |
53 |
50 |
| 6A |
40 |
65 |
65 |
65 |
65 |
60 |
57 |
55 |
| 10A |
33 |
65 |
65 |
65 |
65 |
65 |
55 |
55 |
| 20A |
25 |
65 |
65 |
65 |
65 |
60 |
57 |
45 |
|
|
|
| USA Customer Service: |
1-800-468-2023 |
| Corcom Products Direct: |
1-847-680-7400 |
| Germany: |
49-89-6089-0 |
|
|
Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change. |